Description
Metrohm Autolab PGSTAT101, the affordable potentiostat/galvanostat with measuring maximum of 100mA and compliance voltage of 10V. The PGSTAT101 is a small, lightweight, and affordable potentiostat/ galvanostat that is ideal for most basic electrochemical research applications.
Compliance voltage + 10 V, max current + 100 mA, and offering all standard electrochemical techniques
On-board integrator for coulometric measurements and current integration cyclic voltammetry
Small, lightweight, and affordable
The entry level in the Autolab range of electrochemical instruments, the Autolab PGSTAT101 in combination with the powerful NOVA software, can be used for most of the standard electrochemical techniques. Autolab PGSTAT101 is an affordable potentiostat/galvanostat without compromising on quality and specifications, making it an ideal instrument for students and educational purposes.
The small footprint allows you to place a high quality potentiostat/galvanostat on a crowded workbench. Analog and digital inputs and outputs for interfacing and controlling external devices are available. The Autolab/PGSTAT101 comes with an internal dummy cell and a built-in analog integrator.
Included parts
– Power cable
CABLE.PWR
Standard power cable for Autolab instruments and accessories.
– Standard USB cable
CBL.USB
Standard USB cable for Autolab instruments.
– Cell cable
CELLCBL.M101.204
Standard cell cable, 1.5 m, with connection for counter electrode (CE), reference electrode (RE), sense electrode (S), working electrode (WE) and ground for M101/M204/PGSTAT204.
– Advanced software for electrochemical research
NOVA
NOVA is the package designed to control all the Autolab instruments with USB interface.Designed by electrochemists for electrochemists and integrating over two decades of user experience and the latest .NET software technology, NOVA brings more power and more flexibility to your Autolab potentiostat/
– NOVA.x
NOVA installation USB stick, including NOVA license.
- Tech specs
| Potential range in volt | -10 V to 10 V |
|---|---|
| Compliance voltage range in volt | ± 10 |
| Maximum current in ampère | ± 0.1 |
| Number of current ranges | 7 |
| Number of current ranges remarks | 10 nA to 10 mA |
| Modular instrument | false |
| Maximum number of modules | 0 |
| Analog integrator | true |
| Multichannel instrument | false |
| Potential and current accuracy | V: ± 0.2% ± 2 mV and i: ± 0.2% ± 0.2% of current range |
| Potential resolution | 3 µV (gain 100) |
| Current resolution | 0.0003% of current range |
| Maximum bandwidth in Hz | 1 M |
| Input impedance in Ohm | 100 G |
| Dimensions in mm (W/H/D) | 90/150/210 |
| Dimensions remark | without cables |
| Weight of the instrument in kg | 2.1 |
Manufacture Datasheet & User Manual :